Gamma-neutron Personal Radiation Detector.
The PM1703GNA-II is a gamma-neutron modification equipped with a high-sensitive scintillator for measurement of the personal dose rate up to 300 µSv/h.
The PM1703®-II series of personal radiation detectors are extremely sensitive and robust devices that detect and locate even trace amounts of radioactive materials.
Equipped with a clip for multiple wearing options and easy to operate even for non-specialists, the PRDs became the perfect fit as Radiation Pagers for public security agencies, including border control, rescue teams, police, and anti-terrorist units, and first responders that need to quickly search for radioactive materials before and during public mass events.
Features
- ANSI N42.32-2016 and IEC 62401:2017 compliance
- NORM-suppression algorithm for differentiating color-coded alarms triggered by natural or man-made radiation materials
- Dedicated 0-9 scale mode with unitless dose rate indication allowing for ease of use and minimal training
- Operation in extreme temperatures from –40 °C to 50 °C
- Long-life alkaline or rechargeable battery for 800 hours
- Shockproof hermetic case IP65
- Audible, visual, and vibration alarms
- Free Polismart® iOS and Android app for advanced operation
- USB and Bluetooth (PM1703GNA-II BT) communication
Applications
- Customs and border control
- HAZMAT and CBRNe teams
- Steel and recycling industry
- Waste management sites
- Counter-terrorism teams
- Homeland security
- First responders
- Special forces
- Public safety
Operation principle
While ensuring high-performance detection for any scenario, the PM1703®-II PRD provides confidence in personal safety by continuously monitoring the measured dose equivalent rate and alerting the user with visual, audible, and vibration alarms when the pre-set radiation thresholds are exceeded.
The operating history is stored in the instrument‘s non-volatile memory (up to 2000 data points), protecting the data even when the battery is removed. The stored data can also be transferred to a PC via USB. The storing data format is designed to meet the ANSI N42.42.
Polimaster PRDs comply with the ANSI N42.32-2016 and IEC 62401:2017 standards, can operate in extreme temperatures, have a long-life battery, and offer a significant advantage with their NORM suppression algorithm. The algorithm differentiates alarm signals when there is an increase in natural gamma background values and detection of Naturally Occurring Radioactive Materials (NORM). When NORMs are detected, a green indication light is displayed. However, if other radionuclide types (IND, NUC, MED) are detected, a red indication light and sound alarm are activated. This feature ensures clear and distinct indications based on the specific type of radiation detected.
The Bluetooth module enables communication with smartphones for advanced operation via the free Polismart® App.
Models
The PM1703GNA-II PRD is the basic model.
The PM1703GNA-II BT PRD is the model equipped with a Bluetooth module that enables communication with smartphones for advanced operation via the free Polismart® Android App.
Detector | Csl(TI) SiPM 6LiF/ZnS |
Dose rate measurement range | 0.1 µSv/h to 300 µSv/h |
Dose rate measurement accuracy | ± 30% |
Energy range | 33 keV to 3.0 MeV (search) |
Neutron radiation energy range | thermal to 14 MeV |
Gamma radiation sensitivity | 100 cps per µSv/h (137Cs) 500 cps per µSv/h (241Am) |
Neutron radiation sensitivity | 0.035 cps⋅cm2/neutron (Pu-α-Be) 1.2 cps⋅cm2/neutron (thermal neutrons) |
Response time | 0.25 s |
Alarms | visual, audible, vibration |
Memory | 2000 events |
Smartphone communication | Bluetooth 4.0 (PM1703GNA-II BT) |
PC communication | USB |
Power supply | one AA alkaline or rechargeable battery |
Battery lifetime | 800 hours |
Ingress protection | IP65 |
Drop test | 1.5 m |
Dimensions | 87 × 72 × 32 mm |
Weight | 200 g |
Operating temperature | –40 °C to 50 °C |
Atmospheric pressure | 84 to 106.7 kPa |
Humidity | up to 98 % at +35 °C |
Mean service lifetime | 8 years |